PURELAB® Option-Q

Features
  • Delivers up to 15 liters an hour of type I water
  • Up to 18.2 MΩ-cm, type I water suitable for critical applications
  • Provides water free from Endotoxins, RNase and DNase
  • Consistent water quality is guaranteed at point of use through recirculation of water through integrated purification technologies and reservoir
  • Easy to maintain
  • Produces type I water directly from tap water
  • Full service and validation support
Tech Specs
Product Specifications Option-Q 7 Option-Q 15 with ELGA Biofilter
Inorganics (resistivity at 25°C) Up to 18.2 MΩ-cm Up to 18.2 MΩ-cm Up to 18.2 MΩ-cm
Organics (TOC) 1 – 3 ppb 1 – 3 ppb 1 – 3 ppb
Bacteria <1 CFU/ml <1 CFU/ml <0.1 CFU/ml
Bacterial endotoxin - - <0.001 EU/ml
pH Effectively neutral
Silica 30 ppm 30 ppm 30 ppm
DNase - - <20 pg/ml
RNase - <0.002 ng/ml
Delivery flow rate
Case studies
PURELAB Option Q
Review us on

Acquired the Option Q7 it is a high quality and savings in equipment consumables. Their constructive lines are of the highest level. Easy to use, easy to do Sanitation, anyone can change the cartridge, the technical support of Elga with his technicians in Brazil and very good. It is a device that has two water production nives grade I and II and a tank with adequate capacity for each User.

Consumables
Composite Vent Filter
Composite Vent Filter
Part Number
LC136M2
Life expectancy
12 months
Impurities Removed
Impact of Overuse

Increase in airborne contaminants and potential recontamination of purified water

DI pack render image
DI Pack
Part Number
LC163
Life expectancy
12 months
Impact of Overuse

Weakly bound ions are released back into the water reducing resistivity

POU filter
Point-of-use Filter
Part Number
LC145
Life expectancy
3 months
Impact of Overuse

Potential recontamination of purified water and reduced flow rates

Pre-treatment Pack
Part Number
LC140
Life expectancy
6 months
Impact of Overuse

Chlorine, chloarmine and organics remain, increasing the burden on downstream consumables which will need more frequent and expensive changes

Reverse Osmosis
Reverse Osmosis
Part Number
LC143
Life expectancy
24-36 months
Impact of Overuse

High levels of organics and inorganics remain in the water reducing the functional life of the DI pack

UV Light render image
UV Light
Part Number
LC118
Life expectancy
12-18 months
Impurities Removed
Impact of Overuse

Uncontrolled bacterial contamination and TOC levels are not reduced